ignis DR

ignis DR


Long life & Stable High Performance

Untitled-5
  • Excellent Sensitivity & Resolution
  • Excellent Image Quality
  • Excellent Reliability

Excellent Sensitivity & Resolution

  • Toshiba’s advanced and proven fine structured CsI:Tl and direct vapor deposition technology deliver higher sensitivity and better resolution.
  • Reflection coating on CsI:Tl screen enables excellent Detective Quantum Efficiency (DQE) and low noise.
  • A high Modulation Transfer Function (MTF) FPI is the result of these outstanding technologies diagnostic images, while delivering the benefits of lower radiation dose to the patient. The FDX3543RP achieves a new level of functionality and reliability in developing functional FPI’s for system manufacturers.

Excellent Image Quality

  • Achieves a raw image with a low-noise through the use of Toshiba’ s own circuit technology.
  • Images have a shorter lag time with an improved refresh function, suitable for radiographic double exposure.
  • Driver software (attached to the product) provides excellent, clear raw images, adopting ghost image correction, offset correction, gain correction and defect corrections.

Excellent Reliability

  • Excellent durability by using CsI:Tl screen direct vapor deposition method.
  • The structure is highly reliable and protected from degradation due to the use of a unique moisture-proof sealing method for the CsI:Tl screen.

SPECIFICATIONS


Application General radiopraphy
Technology Quadcel technology
TFT TFT array + Photodiode (a-Si)
Pixel pitch 143μm
Active area 35(H) x 43(V) cm (14 X 17 inches)
Active pixel matrix 2448(H) x 2984(V)
A/D conversion 16 bits
Image output time 3 seconds for full image
Cycle time 6 seconds (min.) / 1 minutes (Std.)
9 seconds (with image lag reduction)
Environment Operation 10 to 35oC
10 to 85% RH (non-condensing)
Storage -15 to 55oC
10 to 90% RH (non-condensing)
Dimensions 384 x 460 x15mm
Weight (approx.) 3kg
Mechanical load 150kg over front panel
100kg on 40 mm dia
Certification IEC60601-1,IEC60601-1-2
MDD93/42/EEC (CE marking)

Application General radiopraphy
Technology Quadcel technology
TFT TFT array + Photodiode (a-Si)
Pixel pitch 143μm
Active area 3008(H) x 43(V) cm (17 X 17 inches)
Active pixel matrix 3008(H) x 3072(V) (approx. 9.2million pixels)
A/D conversion 14 bits
MTF (2 lp/mm) 36% typ
DQE (0) 70% typ
Environment Operation 10 to 35oC
30 to 85% RH (non-condensing)
Storage -15 to 55oC
10 to 90% RH (non-condensing)
Dimensions Sensor Unit : 512 x 495 x43mm
Weight (approx.) Sensor Unit : (16kg or 9kg)
Certification IEC60601-1,IEC60601-1-2
MDD93/42/EEC (CE marking)

ignisDR 5thGeneration X-ray Flat Panel Detector

Quadcel is the name for the four core technologies that improve the performance of FPD products

Untitled-6
A : In – house CsI:TI
Untitled-7

Superstructure ( Top View)

High Performance & Low dose

  • Advanced in-house technology which is cultivated in long history and experience
  • X-ray Scintillator produced by the optimum process control for Flat panel Detector
  • The technology enables high performance & low dose
B : Direct CsI : TI vapor deposit;
Untitled-7-1

Superstructure ( Side View)

High Performance & Low dose

  • Fine CsI :TI Structure on Photodioda / TFT board manufactured by advanced process Control
  • The technology enables high performance & low dose
C : Reflective layer coating
8

High Performance & Low dose

  • 1.8 times sensitivity against Non-Reflective layer coating
  • The technology enables high sensitivity & low dose
D : Moisture-proof sealing
9

Long life & Stable high Performance

  • High reliability is achieved by advanced sealing
  • The technology enables long life & stable high perfomance